| Model Number |
5038-58 |
| Medical Device Problem Codes |
Failure to Capture (1081); Electromagnetic Interference (1194); Over-Sensing (1438); Low impedance (2285)
|
| Health Effect - Clinical Code |
Syncope (1610)
|
| Date of Event |
01/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited a warning and polarity switch due to low impedance values and oversensing.The lead further exhibited no capture and the patient experienced syncopal episodes.It was noted that the oversensing was likely due to electromagnetic interference.The right ventricular (rv) lead was capped and replaced and the ipg was explanted and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was below the expected lower range.Analysis of the device memory indicated oversensing associated with the right ventricular lead.If information is provided in the future, a supplemental report will be issued.
|
| |
|