| Catalog Number |
386807 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd cathena¿ safety iv catheter the anti reflex valve does not function after insertion or connection/disconnection and there was blood leaking past the valve and out of the hub.Insertion was observed on live patient, steps for insertion were performed correctly, and as soon as the needle was pulled out of the hub, the blood would leak up by following the spring inside the catheter hub.Connection with a saline flush was performed, after disconnection, catheter leaked = valve not functioning either.Repeated this twice, with the same result.
|
| |
|
Event or Problem Description
|
|
It was reported that bd cathena¿ safety iv catheter the anti reflex valve does not function after insertion or connection/disconnection and there was blood leaking past the valve and out of the hub.Insertion was observed on live patient, steps for insertion were performed correctly, and as soon as the needle was pulled out of the hub, the blood would leak up by following the spring inside the catheter hub.Connection with a saline flush was performed, after disconnection, catheter leaked = valve not functioning either.Repeated this twice, with the same result.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: 5 representative samples (unopened) were returned for investigation.The samples were subjected to blood escape time (bet) test.The samples passed the bet test, no leakage was observed.A review of past 12 months quality notification was performed and no quality notification was raised on the reported defect of leakage.No abnormality was detected during the production of the reported batch.The returned sample passed the bet test.Hence root cause could not be determined.However, there was a capa raised for blood droplet formation at the luer end of the catheter adapter (refer to capa#86125).Small id catheter tubing and small septum vents have been implemented to enhance the blood control function.
|
| |
|