| Model Number |
5913230 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use on patient, the introducer sheath was allegedly identified to be 15f instead of 16.5f through the packaging.It was further reported that the device was not opened or used.There was no reported patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 16 fr d/l equistream xk hemodialysis x23 cm catheter with surecuff kit was returned for evaluation.Visual and microscopic evaluations were performed.The investigation is inconclusive for incorrect component, as gross examination showed the peal-apart sheath to be 15.0 fr with 5.52 mm od x 13.2 cm with an 17.5 cm introducer.However, due to the kit being opened upon return it is unknown whether the mixed components could have been mixed up in clinical or packaging return procedure.The most likely root cause is due to a mix-up at the supplier where the incorrect sheath was mixed with the correct sheath.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use on patient, the introducer sheath was allegedly identified to be 15f instead of 16.5f through the packaging.It was further reported that the device was not opened or used.There was no reported patient contact.
|
| |
|
Search Alerts/Recalls
|