| Model Number |
MMT-522NAL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had keypad unresponsive.The customer blood glucose level was unknown.Customer stated that the button were not responding.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.
|
| |
|