• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS METRX SYSTEM; LIGHT, SURGICAL, CARRIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS METRX SYSTEM; LIGHT, SURGICAL, CARRIER Back to Search Results
Catalog Number 9560702
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 02/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Although it is unknown if the device caused or contributed to the reported event or not, we are filing this report for notification purposes.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that intra-op, heat burn to patient's back where connection of light source to light cord attaches.Radiance light source used with size 20 tubes.The protective covering was not slid over the light source to light cord connection.After first tlif was complete, burns were noticed on the patients back and identified to be caused by the light source/light cord connection.A vaseline gauze was immediately placed on the burn, and the covering slid over the light source to light cable connection.
 
Additional Manufacturer Narrative
Additional information: d10, h3, h6 h6: product analysis results: there does not appear to be a problem with the light source.Per the reported event the cover was not pulled over the connector.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRX SYSTEM
Common Device Name
LIGHT, SURGICAL, CARRIER
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key8398212
Report Number1030489-2019-00238
Device Sequence Number13531094
Product Code FSZ
UDI-Device Identifier00721902444104
UDI-Public00721902444104
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K993021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/23/2022
Device Catalogue Number9560702
Device Lot Number0601680W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2020
Initial Date Received by Manufacturer 02/08/2019
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date03/07/2019
Supplement Report FDA Received Date07/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-