• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number SYM3420E
Medical Device Problem Codes Migration or Expulsion of Device (1395); Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Emotional Changes (1831); Unspecified Infection (1930); Inflammation (1932); Necrosis (1971); Pain (1994); Scar Tissue (2060); Seroma (2069); Hernia (2240); Injury (2348); Fluid Discharge (2686); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced infection, pain, recurrence, balled up mesh and fat necrosis.Post-operative patient treatment included revision surgery.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of multiple ventral i incisional hernias.It was reported that after implant, the patient experienced flap necrosis, infection, pain, recurrence, balled up mesh and fat necrosis.Post-operative patient treatment included revision surgery, revision of flap following abdominal wall reconstruction, excision of necrotic tissue, fat necrosis removed and one area of redundant mesh was excised.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of multiple ventral i ncisional hernias.It was reported that after implant, the patient experienced mental pain, inflammation, adhesions, scarring, suffering, mesh migration, defective device, mesh failure, flap necrosis, infection, pain, recurrence, balled up mesh and fat necrosis.Post-operative patient treatment included revision surgery, revision of flap following abdominal wall reconstruction, excision of necrotic tissue, fat necrosis removed, mesh revision, and one area of redundant mesh was excised.
 
Additional Manufacturer Narrative
Additional info: b5, e (facility name, street, city, region, postal code), g1, h6 (patient codes and device codes).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of multiple ventral i ncisional hernias.It was reported that after implant, the patient experienced seroma, slough in midline wound, mental pain, inflammation, adhesions, scarring, suffering, mesh migration, defective device, mesh failure, flap necrosis, infection, pain, recurrence, balled up mesh, and fat necrosis.Post-operative patient treatment included medications, revision surgery, revision of flap following abdominal wall reconstruction, excision of necrotic tissue, fat necrosis removed, mesh revision, and one area of redundant mesh was excised.
 
Additional Manufacturer Narrative
Additional info: b5, b7, d8, g1, g4 (pma / 510(k)), h6 (patient and health impact codes).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMBOTEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key8400335
Report Number9615742-2019-00415
Device Sequence Number2326477
Product Code FTL
UDI-Device Identifier10884521190542
UDI-Public10884521190542
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K142908
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberSYM3420E
Device Catalogue NumberSYM3420E
Device Lot NumberPRF0115X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/13/2019
Supplement Date Received by Manufacturer05/15/2019
05/15/2019
10/26/2022
Initial Report FDA Received Date03/07/2019
Supplement Report FDA Received Date10/22/2019
07/26/2022
11/11/2022
Date Device Manufactured06/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight76 KG
-
-