| Model Number |
015-044 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that when the doctor used the matching screw to the plate being used to break into the skull, the plate's hole was broken and could not be used normally.The operation was completed after replacing with the same type of product.The patient's status at the time of the report was alive-no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
One timesh plate was returned.The plate was not attached to the tag with the part number and lot number information.One of the end eyelets was broken and left a gap through the aperture.It is unknown how this damage occurred.The instructions for use cautions, ¿excessive force, over-bending, notching, or scratching during mesh forming process could result in mesh fracture.¿ if information is provided in the future, a supplemental report will be issued.
|
| |
|