| Model Number |
D314TRM |
| Medical Device Problem Codes |
Failure to Charge (1085); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an alert on the cardiac resynchronization therapy defibrillator (crt-d) for a charge circuit timeout.An interrogation report showed the device was at end of service (eos) one month prior to this.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|