| Catalog Number |
04.01.0160 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 07 march 2019: lot 182293: (b)(4) items manufactured and released on 04-sep-2018.Expiration date: 2023-08-23.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs director: screw breakage during implantation.The screw was replaced by a new one and surgery was completed successfully.No clinical consequence is expected for this event.
|
| |
|
Event or Problem Description
|
|
The surgeon may have misplaced the screw during surgery and the head of the screw broke.He decided to use another screw to complete the case.
|
| |
|