• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOLDING SLEEVE 85MM FOR MEDIUM DISTRACTOR; TRACTION,APPARATUS,NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOLDING SLEEVE 85MM FOR MEDIUM DISTRACTOR; TRACTION,APPARATUS,NON-POWERED Back to Search Results
Model Number 395.46
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Occupation: reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during open reduction internal fixation (orif) of ankle surgery on (b)(6) 2019, the wing nut broke on medium distractor sleeve pin and stuck in sleeve.Fragments were generated and removed easily without additional intervention.There was no surgical delay.Procedure was successfully completed.No patient consequence.This report is for one (1) holding sleeve 85mm for medium distractor.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: device history lot part: 395.460, lot: 9319839, manufacturing site: betlach, release to warehouse date: 02.Feb.2015.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The material was reviewed and was confirmed to meet the specification with no non-conformance noted.H3, h6: investigation summary: complaint summary: 03/28/2019: updated event description: it was reported that on (b)(6) 2019, during open reduction internal fixation (orif) of ankle, the wing nut broke on medium distractor sleeve pin and stuck in sleeve.Fragments were generated and removed easily without additional intervention.The schanz screw blunted trocar point was found stripped/worn during an investigation of the device.There was no surgical delay.Procedure was successfully completed.No patient consequence.Flow: broken visual inspection: the holding sleeve 85 mm for medium distractor (part # 395.46, lot # 9319839, mfg # 02-feb-2015) was received at us cq with the wing screw broken off and not returned.This is consistent with the reported complaint condition, thus confirming the complaint.Functional test: a functional test was performed and the schanz screw and the holding sleeve with distractor are stuck together and unable to be disassembled.The complaint was able to be replicated, therefore the complaint is confirmed.Dimensional analysis was not performed as the wing screw was completely broken off and not returned.Document/specification review: the material was reviewed and was confirmed to meet the specification with no non-conformance noted.Conclusion: the complaint condition is confirmed as the holding sleeve 85 mm for medium distractor (part # 395.46, lot # 9319839) was received with the wing screw broken off.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces.No new malfunctions were observed during the course of this investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.D4: lot number provided for reporting.H3, h6: the device was received, the investigation completed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
03/28/2019: updated event description.It was reported that on (b)(6) 2019, during open reduction internal fixation (orif) of ankle, the wing nut broke on medium distractor sleeve pin and stuck in sleeve.Fragments were generated and removed easily without additional intervention.The schanz screw blunted trocar point was found stripped/worn during an investigation of the device.There was no surgical delay.Procedure was successfully completed.No patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLDING SLEEVE 85MM FOR MEDIUM DISTRACTOR
Common Device Name
TRACTION,APPARATUS,NON-POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8404526
Report Number2939274-2019-56881
Device Sequence Number1811050
Product Code HST
UDI-Device Identifier10886982200334
UDI-Public(01)10886982200334
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number395.46
Device Catalogue Number395.46
Device Lot Number9319839
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2019
Initial Date Received by Manufacturer 02/14/2019
Supplement Date Received by Manufacturer03/28/2019
Initial Report FDA Received Date03/08/2019
Supplement Report FDA Received Date04/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-