• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27MJ-501
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Mitral Regurgitation (1964); Thrombus (2101); No Consequences Or Impact To Patient (2199)
Date of Event 08/30/2018
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2018, a 27mm masters valve was implanted in the mitral position.During the three month follow-up, the patient was confirmed to have severe regurgitation which required intervention.On (b)(6) 2018, the device was explanted.The patient didn't remain hemodynamically stable throughout the procedure; the cross-valve differential pressure was high.The patient was administered additional warfarin to stabilize.The physician believes the regurgitation was due to the size of the valve.Per the physician there was thrombus on the device.The device was replaced and a 29mm masters valve was implanted.The patient was stable post-operatively.
 
Additional Manufacturer Narrative
An event that the physician chose the improper valve size and while handling the valve before implant the valve fell to the floor was reported.No damage or other anomalies were found on the valve and there was no evidence of implant, consistent with information from the field.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the mis-sizing and contamination event could not be conclusively determined.
 
Event or Problem Description
It was initially reported that the reported valve was explanted after an implant period of about 3 months.Based on a follow-up conversation with the physician, it was confirmed that this 27mm valve was selected for implant, but determined to be the incorrect size.As the device was exchanged the device fell onto the floor.The procedure was completed with a replacement device.The reported device was never implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8404667
Report Number2648612-2019-00020
Device Sequence Number18937634
Product Code LWQ
UDI-Device Identifier05414734006460
UDI-Public05414734006460
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/18/2023
Device Model Number27MJ-501
Device Lot Number6417419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/13/2019
Supplement Date Received by Manufacturer04/25/2019
Initial Report FDA Received Date03/08/2019
Supplement Report FDA Received Date05/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-