The device was not returned for analysis as it was discarded at the site.Since the device was not returned, we are unable to perform further root cause analysis.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.Per our instruction for use (ifu): "gently immerse the axium detachable coil and its detachment zone in heparinized saline.Take care not to stretch the coil during this procedure, in order to preserve the coil memory.While still immersed in the heparinized saline, point introducer sheath vertically into saline and gently retract the distal tip of the coil into the introducer sheath.In order to achieve optimal performance of the axium detachable coil and to reduce the risk of thromboembolic complication, it is advised that a continuous saline flush be maintained between the femoral sheath and the guiding catheter, the micro catheter and guiding catheter and the micro catheter and the implant delivery pusher and the axium detachable coil.Place the appropriate guiding catheter following recommended procedures.Connect a rotating hemostatic valve (rhv) to the hub of the guiding catheter.Attach a 3-way stopcock to the side arm of the rhv, then connect a line for the continuous flush.Attach a second rhv to the hub of the micro catheter.Attach a 1-way stopcock to the side arm of the rhv, then connect a line for continuous flush.For axium detachable coils: one drop from the pressure bag every 3-5 seconds is suggested.For pgla or nylon microfilament axium detachable coils: one drop from the pressure bag every 1-3 seconds is suggested.Check all fittings so that air is not introduced into guiding catheter or micro catheter during continuous flush.It is essential to confirm catheter compatibility with the axium detachable coil.The outer diameter of the axium detachable coil should be checked to ensure that the coil will not block the catheter.Do not advance the coil with force if the coil becomes lodged within or outside the microcatheter.Determine the cause of resistance and remove the system when necessary.Insert the distal end of the introducer sheath through the rotating hemostatic valve (rhv) and into the hub of the microcatheter until the sheath is firmly seated.Tighten the rhv around the introducer sheath to prevent back flow of blood, but not so tight as to damage the coil during its introduction into the catheter." if information is provided in the future, a supplemental report will be issued.
|