| Model Number |
6996SQ58 |
| Medical Device Problem Code |
Low impedance (2285)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient reported hearing an alert.Low defibrillation impedance was noted on both the svc coil and rv coil of the patient's subcutaneous leads.The coils were too close together which was triggering the alert.The leads were explanted and replaced with 1 subcutaneous lead.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported/clarified that the patient had 2 revision procedures.During the first procedure, one subcutaneous lead was removed and replaced with an intravenous right ventricular (rv) lead.The following day, the 2nd subcutaneous lead was removed and replaced with a new subcutaneous lead.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|