ARTHROCARE CORP. AMBIENT KVAC IFS WAND; ARTHROSCOPE Back to Search Results
Catalog Number ASHA5050-01
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/19/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the kvat wand came off in the middle of a hipscope.There was a lot of pressure and force being applied to the tip at the time of the incident.As is often the case in a hipscope.The item was retrieved successfully and the operation continued.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Factors not associated with the manufacture or design of the device which could result in damage to the tip could be the device coming into abrupt contact with a hard (metallic) object, improper insertion or removal from an apparatus, use at a higher than recommended set point or excessive force applied to the tip as reported by the customer can also lead to unintended detachment.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBIENT KVAC IFS WAND
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8406714
Report Number3006524618-2019-00136
Device Sequence Number2227971
Product Code HRX
UDI-Device Identifier00817470000252
UDI-Public00817470000252
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K162126
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberASHA5050-01
Device Lot Number1156919
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/19/2019
Supplement Date Received by Manufacturer04/04/2019
Initial Report FDA Received Date03/11/2019
Supplement Report FDA Received Date04/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-