| Model Number |
MMT-551NAP |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
02/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump was under delivering and they experienced high blood glucose level.Customer¿s current blood glucose level was 382 mg/dl.The customer feels ok to troubleshooting high blood glucose level.Customer reported that they did not contacted their health care professional regarding the high blood glucose event.The customer did not provide any symptoms regarding high blood glucose.The customer was treated for the high blood glucose with injections.The drive support cap appears normal.The customer was neither in the emergency room, nor admitted into hospital as a result of high blood glucose.Based on customer report customer does allege insulin pump was under delivering.The insulin pump and the reservoir was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during basic occlusion test due to faulty force sensor resistor (gold).Unable to perform basic occlusion test, occlusion test, prime test, excessive no delivery test or verify displayable history crc check failed on startup alarm or possible over/under delivery anomaly due to motor error alarm.Motor passed motor test.Additional information has been received which was not included with the initial report.The information has been provided with this report.
|
| |
|
Event or Problem Description
|
|
It was reported via phone call that the customer's weight has been changed to (b)(6).
|
| |
|