| Model Number |
MMT-723RNAH |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they had trouble during filling the insulin pump.The customer¿s blood glucose was 59 mg/dl.The customer was assisted with troubleshooting.Customer states they were unable to exit the preparing to prime loop.Customer states the drive support cap appears normal or recessed.Customer states they did not receive second series of beeps nor did numbers appear on the screen.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No prime/fill anomaly noted during test.
|
| |
|