| Catalog Number |
MBL-6-XL-C |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
|
| Date of Event |
02/14/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Trigger cord entrapment and difficulty in effectively deploying bands can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a colonoscopy with banding, the physician used a cook 6 shooter saeed multi-band ligator.After the last band was deployed, the string [trigger cord] from the device was stuck under the band.They cut the string and removed the endoscope.They went back in with the endoscope to cut the string at the band [in order to remove the endoscope].They do not know if any of the string remained in the patient.Other than the deployed band, it is unknown if a portion of the string of the device remained inside the patient's body to pass naturally.The patient required the string to be cut from the band in order to remove the endoscope.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|