| Catalog Number |
04P54-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.For the (b)(6) results on this patient, see 3008344661-2019-00026.
|
| |
|
Event or Problem Description
|
|
The customer observed a (b)(6) result on the architect i1000sr analyzer.The following data was provided: initial (b)(6), repeats (b)(6); confirmed (b)(6).The sample was confirmed (b)(6) at another laboratory.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of the design history file, and a review of labeling.No adverse trend, non-conformances, potential non-conformances, or deviations were identified for the customer's issue.No return patient sample was available.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|