• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 3000 ML; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 3000 ML; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX15RE30
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the arterial temperature probe ports were unable to record temperature.  no known impact or consequence to patient.The product not changed out.The procedure was completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 11, 2019.  upon further investigation of the reported event, the following information is new and/or changed: (additional device information - added exp date) (date received by manufacturer) (indication that this is a follow-up report) (follow-up due to additional information) (device evaluated by manufacturer) (device manufacture date) (identification of evaluation codes 4114, 11, 3331, 3259, 4307) method code #1: 4114 - device not returned method code #2: 11 - testing of device from same lot/batch retained by manufacturer method code #3: 3331 - analysis of production records results code: 3259 - improper physical structure conclusions code: 4307 - cause traced to component failure the actual sample was not returned for evaluation.A retention sample from the same product/lot number was tested and found to be functioning properly.An engineering investigation and capa have been initiated to determine a definitive root cause.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLLOW FIBER OXY WITH 3000 ML
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8409557
Report Number1124841-2019-00065
Device Sequence Number312318
Product Code DTZ
UDI-Device Identifier00699753450134
UDI-Public(01)00699753450134
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K151389
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/29/2021
Device Model Number3CX*RX15RE30
Device Catalogue NumberN/A
Device Lot NumberWN12
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/21/2019
Supplement Date Received by Manufacturer05/08/2019
Initial Report FDA Received Date03/11/2019
Supplement Report FDA Received Date05/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-