BAXTER HEALTHCARE - LUND MONITORS PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107493
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an air in blood alarm (return pressure -155) occurred during patient treatment with a prismaflex control unit and a prismaplex set.It was reported the alarm could not be resolved.The return line was inspected and air was visible between the air detector and the return line clamp.The customer decided not to return the blood manually because of the visible air.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not made available and therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - LUND MONITORS
lund skane lan
MDR Report Key8410462
MDR Text Key139585815
Report Number9616026-2019-00011
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/11/2019,04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number107493
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/11/2019
Distributor Facility Aware Date02/14/2019
Event Location Hospital
Date Report to Manufacturer03/11/2019
Date Manufacturer Received04/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAFLEX SET.
-
-