| Model Number |
26430520-1 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Pneumothorax (2012)
|
| Date of Event |
01/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Per our evaluation findings: the manifold has an internal co2 leak.The manifold needs to be rebuilt due to wear and tear.The power supply is faulty and damaged.(solder joints) the power harness needs to be upgraded.The control board is faulty and it needs to be replaced due to wear and tear.The unit was manufactured in 2005 and has been in use for over 14 years; it shipped to customer in october 2005 and this is the first time they have sent it in for service.
|
| |
|
Event or Problem Description
|
|
Allegedly, during a gastric sleeve procedure, the patient developed a pneumothorax.
|
| |
|