| Catalog Number |
5442136 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: only the nose portion of the screw was returned for analysis.The witness marks on the nose and fractured surface are consistent with pressure from the rod while under torsion.This combined pressure both parallel and perpendicular to the axis of the screw caused the material to overload.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with t5 fracture and underwent posterior fusion from t2 to t7 levels.During surgery, the top of the crosslink set screw sheared off.The product came in contact with the patient.The fragment of the set screw still remains in the patient.No patient complications were reported.
|
| |
|