A review of the device history record for the blue or towels pwtb04-stm, lot#20180710-23-sh shows that the product was manufactured on 23rd jul 2018.Based on the supplier¿s investigation, the device history record review did not indicate any exception that would have led to the reported incident.The supplier checked on the retained sample.The average linting data was 208/10 pieces and within limits (=0.38g/10 pieces).No sample was provided for evaluation at this time.The or towel is made of cotton, so lint is born and inevitable.The intended use of or towel is to be used for applying medication to or absorbing small amounts of body fluids from a patient's body surface.Measures to better control and improve linting have been implemented.A suction process was added before products' final folding and operators perform this activity according to standard operation procedure requirements.In the folding process, one cloth bag protects 100 pieces of semi-finished products to avoid linting stuck onto the products during product transfer.Based on the investigation, all linting test data is within the acceptable range.There is no abnormal situation that has happened in production.Therefore, the root cause could not be determined and no further action.The complaint information was shared with the relevant sectors within the manufacturing facility for their awareness and we will continue to monitor for this type of incident.
|
|
The customer reported that during a clot retrieval on a stroke, lint from the blue or towels pwtb04-stm from the neuro embolism pack/san11an10h was found on a stent.The lint did not reach the patient, there was no delay and no injury.No patient demographics or further data was provided.
|