• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PFC KEEL TIB TRAY CEM SZ4; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PFC KEEL TIB TRAY CEM SZ4; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 864182
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 02/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
 
Event or Problem Description
Patient was revised to address pain and loosening of the tibial component at the cement to implant interface.Cement manufacturer is competitor.Surgeon reviewed current x-rays suspected loosening and revision was done.Doi: (b)(6) 2017.Dor: (b)(6) 2019; right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC KEEL TIB TRAY CEM SZ4
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8412969
Report Number1818910-2019-87037
Device Sequence Number18907911
Product Code JWH
UDI-Device Identifier10603295216445
UDI-Public10603295216445
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K961685
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number864182
Device Lot NumberD17010252
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/18/2019
Supplement Date Received by Manufacturer04/11/2019
Initial Report FDA Received Date03/12/2019
Supplement Report FDA Received Date04/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-