| Model Number |
LVZRXT180J |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the full guidewire was returned in segments and analyzed.The analysis indicated that the guidewire was kinked/buckled.The guidewire was broken.The guidewire was unraveled.Visual analysis of the lead indicated damage during use.The analyst noted that the proximal portion of the guidewire was received in dispenser, and the distal portion was broke and stuck in lead.Visual inspection found the guidewire tip was kinked inside the lead tip nose.The visual inspection result leads us to conclude that during the procedure the guidewire was kinked.Kinked wire was obstructed by the lead conductor, and could not remove the guidewire from lead.The guidewire broke possible when force was applied.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implant procedure, the guidewire became stuck inside the lead.The guidewire and lead were removed and a replacement lead was implanted.No patient complications have been reported as a result of this event.
|
| |
|