Expedium screw head broke during procedure.Was surgery delayed due to the reported event? - yes, if yes, number of minutes: - 5, action taken when event occurred? - screw removed and replaced , was procedure successfully completed? - yes, were fragments generated? - yes, if yes, were they removed easily without additional intervention? - yes, patient status/ outcome / consequences - no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: - unknown, is the patient part of a clinical study - no.
|
|
Product complaint # = > (b)(4).Udi: (b)(4).Visual examination found that the threads on the screw¿s tulip head were torn off.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn threads cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|