Model Number MALE STEM
Device Problems Gas Output Problem (1266); Misconnection (1399); Component Missing (2306)
Patient Problem Respiratory Distress (2045)
Event Date 12/20/2018
Event Type  Injury  
Event Description
Pin was missing from the male adapter of the regulator allowing it to be inserted into the nitrous oxide outlet.Nitrous oxide was administered instead of supplemental oxygen during and mri procedure.Patient experienced respiratory distress and was provided breathing support to maintain oxygenation levels.Patient was admitted to the pediatric intensive care unit for further monitoring prior to discharge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OHMEDA
Type of Device
Y FLOW METER
Manufacturer (Section D)
TRI-ANIM
MDR Report Key8413610
MDR Text Key138638601
Report Number8413610
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberMALE STEM
Device Catalogue Number792-8-1321-01
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/01/2019
Distributor Facility Aware Date12/28/2018
Event Location Hospital
Date Report to Manufacturer03/01/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/04/2019
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age3 YR
-
-