| Model Number |
BA28-90/I20-30 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Product Quality Problem (1506); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx abdominal aortic aneurysm stent.Approximately four (4) years post initial procedure, the patient presented emergently with abdominal pain; the subsequent ct revealed a type ib (right side) and type iii (a or b at the bifurcation) endoleaks.In immediate response, the right internal was coiled, and the patient was reportedly in stable condition.The physician elected to treat the patient by relining with a bifurcated afx2, an ovation ix extender, and ovation ix iliac limb devices on 14 feb 2019.There have been no additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A clinical evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and or imaging; requests were made and denied.As such, event determination, off-label conditions, related patient harms, and patient disposition could not be independently assessed.The manufacturing lot review confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place july 2014.
|
| |
|