• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA28-90/I20-30
Medical Device Problem Codes Material Puncture/Hole (1504); Product Quality Problem (1506); Material Integrity Problem (2978)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx abdominal aortic aneurysm stent.Approximately four (4) years post initial procedure, the patient presented emergently with abdominal pain; the subsequent ct revealed a type ib (right side) and type iii (a or b at the bifurcation) endoleaks.In immediate response, the right internal was coiled, and the patient was reportedly in stable condition.The physician elected to treat the patient by relining with a bifurcated afx2, an ovation ix extender, and ovation ix iliac limb devices on 14 feb 2019.There have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
A clinical evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and or imaging; requests were made and denied.As such, event determination, off-label conditions, related patient harms, and patient disposition could not be independently assessed.The manufacturing lot review confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place july 2014.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8416379
Report Number2031527-2019-00121
Device Sequence Number1612490
Product Code MIH
UDI-Device Identifier00818009012630
UDI-Public(01)00818009012630(17)171023
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2017
Device Model NumberBA28-90/I20-30
Device Lot Number1251908-023
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/13/2019
Supplement Date Received by Manufacturer02/13/2019
Initial Report FDA Received Date03/13/2019
Supplement Report FDA Received Date04/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0006-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-VELA INFRARENAL:1178446-010; AFX-VELA INFRARENAL: (B)(4)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-