| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Break (1069); High impedance (1291)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4) review of the device ecog and lead impedance confirmed that the ecog data are suggestive of a potential lead break.
|
| |
|
Event or Problem Description
|
|
A change in the signals for the depth lead connected to port 1 of the rns neurostimulator was noticed on (b)(6) 2017.The treating center reported that the patient had a post-ictal altercation on (b)(6) 2017 in which the patient's head hit the ground.On (b)(6) 2018, detection was disabled on the depth lead connected to port 1.During a routine neurostimulator replacement on (b)(6) 2019, the lead was removed and a new lead was implanted and connected to the neurostimulator without complications.
|
| |
|