• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1520
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during esophagectomy/gastric tube procedure, in the middle of procedure about at the tenth sealing, the device sealing became insufficient or half cauterized and it has end tone.There was no bleeding occurred.Replaced with new similar device and it worked fine which completed the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8417642
Report Number1717344-2019-00326
Device Sequence Number1839873
Product Code GEI
UDI-Device Identifier10884521125520
UDI-Public10884521125520
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092879
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberLF1520
Device Catalogue NumberLF1520
Device Lot Number60500050X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2019
Initial Date Received by Manufacturer 02/18/2019
Supplement Date Received by Manufacturer03/25/2019
Initial Report FDA Received Date03/13/2019
Supplement Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-