• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; RSP HUMERAL SOCKET INSERT, 40MM +4MM, STANDARD HXE-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; RSP HUMERAL SOCKET INSERT, 40MM +4MM, STANDARD HXE-PLUS Back to Search Results
Model Number 509-00-440
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/15/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient becoming infected.All implants were removed and a cement spacer was placed until infection was resolved.The patient was reimplanted as intended.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 5.3 months apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
RSP HUMERAL SOCKET INSERT, 40MM +4MM, STANDARD HXE-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8418181
Report Number1644408-2019-00187
Device Sequence Number1303783
Product Code KWS
UDI-Device Identifier00888912144599
UDI-Public(01)00888912144599
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/11/2022
Device Model Number509-00-440
Device Catalogue Number509-00-440
Device Lot Number387P1025
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/15/2019
Supplement Date Received by Manufacturer04/04/2019
Initial Report FDA Received Date03/13/2019
Supplement Report FDA Received Date04/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-