| Model Number |
E360 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The medtronic field service engineer evaluated the device and replaced the loud speakers.The ventilator passed all testing per manufacturing specification and was returned to the customer.If the replaced part is returned for failure investigation, a supplement medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during servicing, the e360 ventilator failed calibration and found the device had no audible alarm with the faulty loudspeakers.The ventilator was not in use on a patient at the time of the reported event. .
|
| |
|