JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MULTIFOCAL IOLS Back to Search Results
Model Number ZMA00
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Vitrectomy (2643); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between 10/12/2018 and 02/04/2019.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens was explanted because it subluxed post-operatively.Reportedly, there was a vitrectomy performed, but no incision enlargement required.The replacement lens was the same model, but a higher diopter (20.5).There was no patient injury reported and the patient is noted to be fine post-op.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-(b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8419030
Report Number9614546-2019-00223
Device Sequence Number16547295
Product Code MFK
UDI-Device Identifier05050474536753
UDI-Public(01)05050474536753(17)200408
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/10/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2020
Device Model NumberZMA00
Device Catalogue NumberZMA00U0175
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/13/2019
Supplement Date Received by Manufacturer03/27/2019
10/25/2020
Initial Report FDA Received Date03/13/2019
Supplement Report FDA Received Date03/28/2019
11/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-