| Model Number |
MMT-723RNAH |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
03/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error during bolus.The customer blood glucose level was unknown.The customer was assisted with troubleshooting.Customer reports the drive support cap appears normal.Customer stated that the insulin pump did not exposed to high magnetic field.Customer was able to successfully clear the alarm.Customer was able to complete insulin pump rewind.Customer stated that the insulin alarm history contains more than one motor error alarm within the last 30 days.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test and displacement test or verify un response button due to blank display.Moisture damage keypad traces during visual inspection.
|
| |
|