• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAH
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 03/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had motor error during bolus.The customer blood glucose level was unknown.The customer was assisted with troubleshooting.Customer reports the drive support cap appears normal.Customer stated that the insulin pump did not exposed to high magnetic field.Customer was able to successfully clear the alarm.Customer was able to complete insulin pump rewind.Customer stated that the insulin alarm history contains more than one motor error alarm within the last 30 days.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test and displacement test or verify un response button due to blank display.Moisture damage keypad traces during visual inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8419336
Report Number3004209178-2019-67372
Device Sequence Number2051065
Product Code OYC
UDI-Device Identifier00643169476578
UDI-Public(01)00643169476578
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAH
Device Catalogue NumberMMT-723RNAH
Device Lot NumberA0723RNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Initial Date Received by Manufacturer 03/08/2019
Supplement Date Received by Manufacturer04/12/2019
Initial Report FDA Received Date03/14/2019
Supplement Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age27 YR
Patient Weight130
-
-