| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump would not exit the preparing to prime loop and insulin squirted out if the tubing during manual prime.The customer¿s blood glucose was 135 mg/dl at the time of the incident.Customer reported insulin squirting out during the prime process.The customer indicated the drive support cap was normal.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test, rewind, basic occlusion, occlusion, prime and excessive no delivery test.Pump primed properly.
|
| |
|