• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES BI-PHASIC
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code Death (1802)
Date of Event 02/15/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.Please refer to the user medwatch report that zoll medical has received.
 
Event or Problem Description
Complainant alleged that while attempting to defibrillate a (b)(6) male patient, the device displayed a "defib pad short" message.Complainant indicated that the clinician obtained another device and electrode pads to continue treating the patient and received the same error.Complainant indicated that the patient subsequently expired.Please reference medwatch report 1220908-2019-00628 for the other defibrillator used in the event.
 
Additional Manufacturer Narrative
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing which included bench handling, multiple 30 joules self-tests, shock testing at various joules settings utilizing the returned multifunction cable and continuity testing without duplicating the reported malfunction.The device's biphasic cable was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.Please refer to the attached user medwatch report that zoll medical has received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8420089
Report Number1220908-2019-00648
Device Sequence Number11275592
Product Code DPS
UDI-Device Identifier00847946013813
UDI-Public00847946013813
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K990762
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/20/2019
Report Date (Section F) 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/15/2019
Event Location Hospital
Date Report to Manufacturer02/19/2019
Initial Date Received by Manufacturer 02/20/2019
Supplement Date Received by Manufacturer02/20/2019
02/20/2019
Initial Report FDA Received Date03/14/2019
Supplement Report FDA Received Date03/29/2019
04/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age39 YR
Patient Weight83
Date Report Sent to FDA02/20/2019
-
-