| Model Number |
K172 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Apnea (1720)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this patient thinks his heart rate is driven up when his continuous positive airway pressure (cpap) is in use.A boston scientific technical services consultant discussed the clinical observations and precautions with this patient.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.The device was exposed to simulated heart load conditions, and the pacing and sensing functions were tested.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.This device operated as expected during laboratory analysis and the associated investigation determined that this device appeared to exhibit an increase in the pacing rate as a result of the minute ventilation (mv) sensor response to instances where the minute ventilation (mv) sensor did not respond as expected at higher rates.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient thinks his heart rate is driven up when his continuous positive airway pressure (cpap) is in use.A boston scientific technical services consultant discussed the clinical observations and precautions with this patient.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.It was reported that this device was subsequently explanted due to normal battery depletion.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|