• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION CATH F4 INF PIG 110CM 6SH; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION CATH F4 INF PIG 110CM 6SH; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 538450S
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint conclusion: during an unknown procedure, when advancing the guidewire into a 4f infiniti (110cm 6sh) pigtail catheter, the pigtail end broke off.There was no reported patient injury.Multiple attempts to gather additional information were unsuccessful.The device was not returned for analysis.A product history record (phr) review of lot 17581813 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The phr review does not suggest that the event reported by the customer could be related to the manufacturing process.The reported ¿ brite tip/distal tip ¿ catheters separated¿ could not be confirmed as the device was not returned for analysis and procedural films/device photographs were not received.It is difficult to draw a clinical conclusion between the device and the event based on the limited information available.However, procedural and handling factors may have contributed to the reported event.Per the instructions for use, which is not intended as a mitigation, ¿exposure to temperatures above 54c (130f) may damage the catheter.To prevent damage to the catheter tip during removal from the package, grasp the hub and withdraw the catheter.Exercise care when removing guidewires from multiple-curve catheters.To prevent kinking of 5f (1.65 mm) and smaller angiographic catheters, and specifically the 4f (1.35 mm) infiniti® pigtail catheters: straighten the pigtail catheter tip only with a diagnostic guidewire or, if applicable, with a tip straightener.Do not straighten by hand.Use a guidewire when introducing the catheter through the catheter sheath introducer (csi) and into the left ventricle.Treat all 4f (1.35 mm) catheters and smaller french sizes with ultimate care.The performance of these products may be impaired if not properly and cautiously handled during unpacking and preparation.¿ neither the phr review nor the available information suggest that the reported event could be related to the manufacturing process.Therefore, no corrective or preventive actions will be taken.
 
Event or Problem Description
As reported, when advancing the guidewire into a 4f infiniti (110cm 6sh) pigtail catheter, the pigtail end broke off.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH F4 INF PIG 110CM 6SH
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key8421456
Report Number9616099-2019-02781
Device Sequence Number2327539
Product Code DQO
UDI-Device Identifier20705032014820
UDI-Public20705032014820
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K862244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Model Number538450S
Device Catalogue Number538450S
Device Lot Number17581813
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/12/2019
Initial Report FDA Received Date03/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-