COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52277
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, an (b)(6) patient (gender unknown) was undergoing an intervention of the superficial femoral artery utilizing a advance 35 lp low profile balloon catheter.The balloon burst circumferentially upon initial inflation in the heavily calcified lesion below nominal pressure.An inflation device was used.The balloon was inflated to 4 atmospheres (atm) once for 3 to 5 seconds.Omnipaque contrast in a 50/50 mixture with saline was used for the inflation.After bursting, the balloon was deflated and removed through a sheath.The lesion was described as 90 to 95 percent occluded.There was no vessel angulation or tortuosity.Another balloon was used to complete the procedure successfully.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation.Reviews of the dimensional verification, complaint history, device history record, drawing, instructions for use (ifu), manufacturer¿s instructions, quality control, specifications, and a functional test & visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned package confirmed that one ptas-35-135-7-14.0 was returned for investigation.Biomatter was present throughout the device.The balloon was inflated with water and a pinhole leak was noted near the middle of the balloon, approximately 6.0cm from the proximal bond.The balloon was then inflated with air in a water filled beaker to photograph the leak.The pinhole in the balloon material was visible as well.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed one nonconforming event which could potentially contribute to this failure mode.This nonconformance was for proximal bond damage.While this is potentially related, the affected device was scrapped and not replaced prior to order completion.It should also be noted there were no other reported complaints for this lot number.Furthermore, reviews of the manufacturer¿s instructions, drawings, quality control procedures, and overall device history file were conducted, and no gaps were discovered.Through this review, cook has determined that appropriate measures are in place to prevent this failure.Moreover, an ifu is provided with the device, which states ¿for percutaneous transluminal angioplasty (pta) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral, and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae." it goes on to say, ¿if balloon pressure is lost and/or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit." based on the information provided and the examination of the returned product, investigation has concluded that a root cause could not be established.Measures are being conducted to address this failure mode.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8421801
Report Number1820334-2019-00565
Device Sequence Number1796266
Product Code DQY
UDI-Device Identifier10827002522777
UDI-Public(01)10827002522777(17)200131(10)7635897
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Model NumberG52277
Device Catalogue NumberPTA5-35-135-7-14.0
Device Lot Number7635897
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2019
Initial Date Received by Manufacturer 02/28/2019
Supplement Date Received by Manufacturer04/16/2019
Initial Report FDA Received Date03/14/2019
Supplement Report FDA Received Date04/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age87 YR
-
-