Model Number 500AHCT
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Event or Problem Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the user saw and smelled smoke from the unit when it was turned on.The unit was turned off and not used throughout the case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed there to be evidence of overheating inside of the hematocrit saturation (hsat) module.The hsat servo card was disassembled and there was a damaged capacitor that appeared to have been overheated.He also noted that he could smell possible overheating had occurred.The service repair technician (srt) installed a new hsat servo card and monitor successfully passed power on self test without the smell of smoke.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Common Device Name
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8422039
Report Number1828100-2019-00127
Device Sequence Number1786128
Product Code DRY
UDI-Device Identifier00886799001622
UDI-Public(01)00886799001622(11)121001
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K133658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/20/2019
Supplement Date Received by Manufacturer04/24/2019
Initial Report FDA Received Date03/14/2019
Supplement Report FDA Received Date05/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-