DEPUY SPINE INC XPDM THORACIC PEDICLE PRB, CRV; PROBE Back to Search Results
Model Number 279702040
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the tip of the pedicle probe broke.It is a old instrument.It happened during surgery, no harm to patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Visual examination at the macroscopic level revealed that the fracture was located at 40mm from the probe¿s tip.The second half of the probe¿s tip was also returned.The fracture analysis report reveals uniform rough morphology of the fractured surface indicating that the instrument underwent a static failure.The absence of rotational deformation and a termination site implies that the probe was broken under a cantilever force.The static overloading led to a fracture of the tip.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the probe¿s distal tip fracturing cannot be positively determined.However, the fracture analysis report reveals uniform rough morphology of the fractured surface indicating that the instrument underwent a static failure due to a single, sudden, unexpectedly high force placed on the tip of the device.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPDM THORACIC PEDICLE PRB, CRV
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8422211
Report Number1526439-2019-51466
Device Sequence Number11208953
Product Code HXB
UDI-Device Identifier10705034198446
UDI-Public(01)10705034198446
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number279702040
Device Catalogue Number279702040
Device Lot NumberNW106670
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2019
Initial Date Received by Manufacturer 02/26/2019
Supplement Date Received by Manufacturer05/16/2019
Initial Report FDA Received Date03/14/2019
Supplement Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-