• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE; VARICOSE VEIN ALBATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE; VARICOSE VEIN ALBATION SYSTEM Back to Search Results
Catalog Number 7201387F
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Information (3190)
Date of Event 02/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have not received the device for evaluation.Hence, we could not conclusively determine the root cause of the defect.The contact person at the hospital stated that the device was picked up by (b)(6).However, the tracking information for this device shows that this unit has not been received by (b)(6).Prior to the procedure, after plugging the handpiece connector into the control unit, the indicator light on the control unit began flashing orange light indicating an issue with the handpiece and the handpiece was found to be non-operational.There has been no serious injury nor the malfunction would result in a death or serious injury if it was to reoccur since the handpiece could not be used for the procedure.However, we have decided to report the incident since our evaluation was based on the reported defect from our sales rep.And the user at the hospital rather than our hands-on evaluation with this defective device itself.The issue was detected during pre-use check.Handpiece was not used in the patient.The procedure was completed using a different handpiece.
 
Event or Problem Description
Handpiece did not operate during pre-use check.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIVEX SYSTEM RESECTOR HANDPIECE
Common Device Name
VARICOSE VEIN ALBATION SYSTEM
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 02148
MDR Report Key8422451
Report Number1220948-2019-00026
Device Sequence Number1795118
Product Code DWQ
UDI-Device Identifier00840663106561
UDI-Public00840663106561
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K032387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/17/2017
Device Catalogue Number7201387F
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2019
Initial Report FDA Received Date03/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-