| Catalog Number |
140.687 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Tissue Damage (2104)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.Moreover, the dentist reported that the implant was immediately installed in an alveolus with signs of injury/ infection.
|
| |
|
Event or Problem Description
|
|
(b)(4) ¿ the dentist reported that 5 months after the dental implant was installed in intraoral region 4#, its non-osseointegration was observed.Moreover, 35n.Cm of primary stability was achieved, the patient presented bone type ii, bone graft was placed and immediate implant procedure was performed.
|
| |
|