| Catalog Number |
168-420 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: broken plug strap and opening on anterior.Leak test and microscopic analysis was performed which identified: opening striated, opening striated puncture and delamination on plug strap (<75% partial separation).The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening due to surgical damage consist in the use of some surgical tool.A striated punctured due to surgical damage like consist in the use of some surgical tool.A sharp broken on plug strap due to an unidentified (tear) opening.A delamination total of the valve (cohesive failure).Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device has been explanted.
|
| |
|