| Model Number |
MMT-551PNAB |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump received with reservoir tube lip completely broken off and missing reservoir tube lip o-ring noted.The test reservoir does not stay locked in place due to reservoir tube lip broken off.Unable to perform displacement test due to reservoir lip broken off.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that their insulin pump was damaged.The customer¿s blood glucose level was 105 mg/dl at the time of the incident.Piece missing from the reservoir compartment.The reservoir was able to lock in place when inserted into pump.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device was expected to be returned for analysis.
|
| |
|