| Catalog Number |
72201395 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Foreign zip code: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, drill tip broke and remained in patients bone as unable to retrieve.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient age between 20 and 30 years.
|
| |
|
Event or Problem Description
|
|
It was reported that during hip arthroscopy labral repair, drill tip broke and remained in patients bone as unable to retrieve.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One 72201395 spade drill bit for 2.9 anchor returned.The complaint stated: ¿drill tip broke during use, drill remained in patients bone as unable to retrieve.¿ the drill bit was returned with the distal tip severed.The distal piece was not returned.The fracture location indicates forceful twist and over torque factors.The condition is consistent with the rotations seizing and therefore the instrument snapped.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.There are no other complaints for this lot.At this time the allegation is considered an isolated incident.
|
| |
|