| Model Number |
PFTS8.0V-NC |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
01/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neomed became aware of user facility report # (b)(4) on 21 january 2019.Additional information was received 13 february 2019 which deemed this complaint reportable.User facility reported an allegation for the potentiality of an adverse event stating the reported stiffened characteristic observed in identified units could cause "damage to the fragile mucosa of a preterm baby's stomach or esophagus".Facility confirmed no adverse event had occurred.Neomed is in the process of working with the facility to obtain more information as well as our medical advisor.Complaint samples have been returned to neomed and evaluations are ongoing.
|
| |
|
Event or Problem Description
|
|
When removing an enteral pvc feeding tube from the patient it was noted that the distal section of tubing had stiffened and was difficult to bend as compared to the section of the tubing out of the neonates body and as compared with a new unused feeding tube in the package.The clinical team also found an additional pvc tube removed later from another patient with this same stiffening characteristic.
|
| |
|
Additional Manufacturer Narrative
|
|
No additional facilities have reported complaints identifying tube stiffness.The underlying root cause relevant to the reported event can not be conclusively determined.
|
| |
|