DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2019, that on (b)(6) 2019, an unexpected receiver shutdown occured.No additional event or patient information is available.No product or data was provided for evaluation.The complaint confirmation of the reported unexpected receiver shutdown could not be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.The device was externally visually inspected and no defects were found.The receiver failed to charge and reboot due to the receiver not turning on.Receiver log download and functional testing were unable to be performed due to the receiver not turning on.The receiver case was opened, the internal inspection failed due to moisture damage.Confirmation of the allegation and a root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ying chen
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8426223
MDR Text Key138994867
Report Number3004753838-2019-26286
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-1
Device Catalogue NumberSTK-GL-010
Device Lot Number5223612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2019
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/18/2019
Initial Date FDA Received03/15/2019
Supplement Dates Manufacturer Received07/05/2019
Supplement Dates FDA Received07/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
Patient Weight61
-
-