| Model Number |
E SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The device was put through extensive testing without duplicating the malfunction.The system to battery interconnect cable and interlock connector was replaced as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib fault 72" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|