| Catalog Number |
TL60 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Device evaluation summary: the analysis results showed that the tl60 device was received with the hook shroud opened by the seam and with a cartridge loaded on the device.The cartridge was returned fully loaded with staples.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the device was noted to have the lockout slide tip and the adjusting knob damaged.The damage to the lockout slid tip and knob is consistent with an excessive force applied to the rotating knob when trying to dial down a locked device, causing the damage to the knob.It should be noted that the instrument has been designed with a lockout feature, which during the first application, prevents turning the adjusting knob unless the retaining pin is pushed completely forward.Please reference the instructions for use for additional information.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported: damaged device.During an unknown surgery, before used on the patient, rotating the knob, noted the plastic shell of handle was damaged.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
| |
|